FDA Adverse Event Malfunction Summary report: N

STERLING BALLOON DILATATION CATHETER

MDR report key: 1944261 · Received January 3, 2011

Report

Report Number
2134265-2010-05795
Event Type
Malfunction
Date Received
January 3, 2011
Date of Event
December 13, 2010
Report Date
December 15, 2010
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053118
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE WAS NOT RECEIVED AT THE COMPLAINT INVESTIGATION SITE FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL / PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE 99% STENOSED LESION WAS LOCATED IN THE MODERATELY TORTUOUS LEFT SUPERFICIAL FEMORAL ARTERY. A 6FR 10CM INTRODUCER SHEATH WAS INSERTED AND A 0.014 175CM GUIDE WIRE WAS ADVANCED. NEXT THE PHYSICIAN ADVANCED THE 5MM X 40MM X 135CM STERLING BALLOON CATHETER ACROSS THE LESION, INFLATED THE BALLOON ONCE TO 6 ATMS, AND THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING BALLOON DILATATION CATHETER CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939031504010 13705842

Patients

Seq Age Sex Outcome Treatment
1 INTRODUCER SHEATH: 6F 10CM TERUMO| INFLATION DEVICE: MEDTRONIC| GUIDE WIRE: 0.014 CRUISE 175CM SJM