LIGAMAX- CLIP APPLIER
Report
- Report Number
- 3005075853-2011-00025
- Event Type
- Malfunction
- Date Received
- January 3, 2011
- Date of Event
- November 19, 2010
- Report Date
- November 22, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT HAD BEEN FIRED THROUGH. PLEASE NOTE THE DEVICE CONTAINS A LOCKOUT FEATURE THAT IS DESIGNED TO INCREASE THE REQUIRED FORCE IT TAKES TO CLOSE THE TRIGGER ONCE THE LAST CLIP HAS BEEN FIRED; THIS REDUCES THE POSSIBILITY OF EMPTY JAWS BEING CLOSED ON A STRUCTURE. IN ADDITION, IF THE TRIGGER IS FORCED CLOSED ONCE THE LOCKOUT HAS BEEN ENGAGED, THE JAWS MAY REMAIN IN THE CLOSED POSITION. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TESTING COULD BE PERFORMED TO EVALUATE THE REPORTED INCIDENT. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIP WAS NOT FED INTO THE JAW PROPERLY. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. THE CUSTOMER DISPOSED OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX- CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |