FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60

MDR report key: 1944214 · Received January 3, 2011

Report

Report Number
1527736-2011-00002
Event Type
Malfunction
Date Received
January 3, 2011
Date of Event
November 17, 2010
Report Date
November 18, 2010
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ARTICULATION LINK. THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE WORKED AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC SLEEVE PROCEDURE, WHEN THE DEVICE WAS CYCLED TO BE FIRED, SOMETHING SOUNDED AND FELT DIFFERENT. THEY STOPPED USING IT, AND OPENED A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 STAPLE, IMPLANTABLE GDW ETHICON ENDO SURGERY, INC (CINCINNATI) UNK UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 (B)(4)