FDA Adverse Event
Malfunction
Summary report: N
ECHELON*FLEX60
MDR report key: 1944214
·
Received January 3, 2011
Report
- Report Number
- 1527736-2011-00002
- Event Type
- Malfunction
- Date Received
- January 3, 2011
- Date of Event
- November 17, 2010
- Report Date
- November 18, 2010
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). ARTICULATION LINK. THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE WORKED AS INTENDED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC SLEEVE PROCEDURE, WHEN THE DEVICE WAS CYCLED TO BE FIRED, SOMETHING SOUNDED AND FELT DIFFERENT. THEY STOPPED USING IT, AND OPENED A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON*FLEX60 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO SURGERY, INC (CINCINNATI) | UNK | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) |