FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER NEW 8

MDR report key: 1944196 · Received December 6, 2010

Report

Report Number
2921482-2010-00950
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
November 11, 2010
Report Date
November 16, 2010
Manufacturer
HOSPIRA, INC.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THE PUMP WAS RETURNED TO THE BIOMEDICAL ENGINEERING DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "MALFUNCTION". NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLUM A+ DRIVER NEW 8 80FRN FRN HOSPIRA, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK LIST #20791, (B)(4)| PLUM A+ SOFTWARE MODULE,