FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 1943782 · Received December 21, 2010

Report

Report Number
2953200-2010-02593
Event Type
Injury
Date Received
December 21, 2010
Date of Event
November 22, 2010
Report Date
November 22, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: ENDOLEAK. SHORT AORTIC NECK. CONVERTER USED AS MAIN DEVICE. EVAL, CONCLUSION: SHORT AORTIC NECK. CONVERTER USED AS MAIN DEVICE.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.4 CM ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK HAD 30 DEGREE ANGULATION, IT WAS 16 MM IN DIAMETER; HOWEVER, ON IVUS IT WAS LARGER THAN 18 MM AND IT WAS 8 MM IN LENGTH. THE ILIAC ARTERIES HAD MILD TO MODERATE CALCIFICATION AND MILD TORTUOSITY. IT WAS REPORTED THAT THE FINAL ANGIOGRAM REVEALED A PROXIMAL TYPE 1 ENDOLEAK. AN ANEURX AORTIC CUFF WAS IMPLANTED, WHICH RESOLVED THE ENDOLEAK. ON (B)(6) 2010 THE PT PRESENTED WITH ABDOMINAL AND BACK PAIN. A CT WAS PERFORMED AND REVEALED THAT THERE WAS THROMBOSIS THROUGHOUT THE ENTIRE STENT GRAFT SYSTEM AND THERE WAS NO BLOOD FLOW THROUGH THE STENT GRAFT (SEE MFR # 2953200-2010-02595 AND 2953200-2010-02594); HOWEVER, THERE WAS SOME COLLATERAL FLOW FEEDING THE LOWER EXTREMITIES. A VASCULAR BYPASS WAS PERFORMED ON THE SAME WEEK THE PT PRESENTED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00524010

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention