TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2010-02594
- Event Type
- Injury
- Date Received
- December 21, 2010
- Date of Event
- December 3, 2010
- Report Date
- December 3, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL, RESULTS: OCCLUSION. MILD TORTUOUSITY, MILD TO MODERATE CALCIFICATION. EVAL, CONCLUSION: MILD TORTUOUSITY, MILD TO MODERATE CALCIFICATION.
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.4 CM ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK HAD 30 DEGREE ANGULATION, IT WAS 16 MM IN DIAMETER; HOWEVER, ON IVUS IT WAS LARGER THAN 18 MM AND IT WAS 8 MM IN LENGTH. THE ILIAC ARTERIES HAD MILD TO MODERATE CALCIFICATION AND MILD TORTUOSITY. IT WAS REPORTED THAT THE FINAL ANGIOGRAM REVEALED A PROXIMAL TYPE 1 ENDOLEAK (SEE MFR # 2953200-2010-02593). AN ANEURX AORTIC CUFF WAS IMPLANTED, WHICH RESOLVED THE ENDOLEAK. ON (B)(6) 2010 THE PT PRESENTED WITH ABDOMINAL AND BACK PAIN. A CT WAS PERFORMED AND REVEALED THAT THERE WAS THROMBOSIS THROUGHOUT THE ENTIRE STENT GRAFT SYSTEM AND THERE WAS NO BLOOD FLOW THROUGH THE STENT GRAFT (SEE MFR # 2953200-2010-02595); HOWEVER, THERE WAS SOME COLLATERAL FLOW FEEDING THE LOWER EXTREMITIES. A VASCULAR BYPASS WAS PERFORMED ON THE SAME WEEK THE PT PRESENTED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00547688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention |