FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 1943776 · Received January 3, 2011

Report

Report Number
2134265-2010-05641
Event Type
Malfunction
Date Received
January 3, 2011
Date of Event
December 7, 2010
Report Date
December 7, 2010
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT WAS NOT RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE A BALLOON RUPTURE OCCURRED. THE 90% STENOSED DE NOVO LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND CALCIFIED DISTAL RIGHT CORONARY ARTERY (RCA). A 12MM X 3.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED TO THE LESION AND UPON INFLATION RUPTURED AT 16ATMS. THE NUMBER OF INFLATIONS AND TO WHAT ATMS IS UNKNOWN. THE BALLOON CATHETER WAS REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912412350 13748995

Patients

Seq Age Sex Outcome Treatment
1 INFLATION DEVICE, ENCORE