FDA Adverse Event Injury Summary report: N

TALENT CAPTIVIA STENT GRAFT SYSTEM

MDR report key: 1943761 · Received December 21, 2010

Report

Report Number
2953200-2010-02591
Event Type
Injury
Date Received
December 21, 2010
Date of Event
November 23, 2010
Report Date
November 23, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070007
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL RESULTS AND CONCLUSIONS: (TORTUOUS AORTA, THORACO-ABDOMINAL ANEURYSM), (ARTERIAL TRAUMA/DISSECTION/PERFORATION), (STENT GRAFT WAS FENESTRATED, AGGRESSIVE BALLOONING INSIDE THE CROWN OF THE GRAFT).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 7 CM THORACO-ABDOMINAL AORTIC ANEURYSM WHICH EXTENDED FROM THE MID-DESCENDING AORTA TO JUST ABOVE THE SMA. THE PROXIMAL AORTIC NECK WAS 31 MM IN DIAMETER AND ENLARGED TO ABOUT 38 MM IN DIAMETER. THE AORTA WAS TORTUOUS. THE PT WAS NOT A CANDIDATE FOR OPEN REPAIR AND HAD A PREVIOUS ABDOMINAL EVAR REPAIR. IT WAS REPORTED THAT TALENT STENT GRAFT WAS PLACED IN THE MID-DESCENDING AORTA AFTER THE PHYSICIAN HAD MADE GRAFT FENESTRATIONS FOR THE CELIAC ARTERY AND OTHER VESSELS. THE STENT GRAFT WAS BALLOONED WITH A RELIANT BALLOON, WHICH WAS PARTIALLY INSIDE THE CROWN OF THE GRAFT. THE PHYSICIAN WAS BALLOONING AGGRESSIVELY AND OVER-INFLATED THE BALLOON. AS A RESULT THE AORTA RUPTURED (SEE MFR # 2953200-2010-02591 AND MFR# 2953200-2010-02592). A 46 MM TALENT THORACIC STENT GRAFT WAS PLACED TO SUCCESSFULLY COVER THE RUPTURE. THE PT DID NOT FEEL WELL AND IS PARAPLEGIC. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT CAPTIVIA STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00565588

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention