FDA Adverse Event Other Summary report: N

EASYPUMP

MDR report key: 1943589 · Received December 8, 2010

Report

Report Number
2026095-2010-00298
Event Type
Other
Date Received
December 8, 2010
Report Date
November 9, 2010
Manufacturer
I-FLOW CORP.
Product Code
MEB
PMA / PMN Number
K052117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO SAMPLE WAS RECEIVED FOR THE EVAL AND INVESTIGATION. WITHOUT THE ACTUAL PRODUCT, A COMPLETE ANALYSIS CANNOT BE CONDUCTED. NO DETERMINATION CAN BE MADE AS TO WHY THE PUMP APPEARED TO INFUSE QUICKLY. THE DIRECTIONS FOR USE (DFU) (1303046, REV. G) TABLE 1 NOTES THAT THE MINIMUM FILL VOLUME FOR THIS PUMP IS 150ML. THE REPORTED FILL VOLUME WAS 255ML SHOULD FLOW APPROX 48 HOURS. THE DFU ALSO CONTAINS A CAUTION STATEMENT: FILLING THE PUMP LESS THAN NOMINAL RESULTS IN FASTER FLOW RATE. NO ADVERSE EVENT OCCURRED. IN ADDITION, THE DFU STATES THAT THE DELIVERY TIMES ARE APPROXIMATELY, AND HAVE AN ACCURACY OF +/- 15% WHEN PUMP IS FILLED TO NOMINAL. THE DEVICE HISTORY RECORD FOR THE ABOVE LOT WAS REVIEWED. ALL MFG OPERATIONS WERE FOUND TO BE WITHIN SPECIFICATION. A REVIEW OF THE LOT HISTORY FOUND NO CONFIRMED COMPLAINTS FOR FAST FLOW FOR THE REPORTED LOT. IF ADDITIONAL INFO PERTINENT TO THIS COMPLAINT OR THE SAMPLE IS RECEIVED, THE FILE WILL BE REOPENED.

Description of Event or Problem · 1

FAST FLOW. CUSOTMER DID NOT WAIT THE 4 HOURS AFTER FILLING BEFORE USE, AS REQUIRED BY THE DIRECTIONS FOR USE. PT EXPERIENCED A HEADACHE AND DIZZINESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYPUMP ELASTOMERIC PUMP MEB I-FLOW CORP. ST 400-4 012451

Patients

Seq Age Sex Outcome Treatment
1