FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 1943495 · Received January 3, 2011

Report

Report Number
2939301-2010-11235
Event Type
Malfunction
Date Received
January 3, 2011
Report Date
December 24, 2010
Manufacturer
LIFESCAN INC.
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K)# IS K073231.

Additional Manufacturer Narrative · 1

FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT 01/24/2011. THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS:THE METER INVOLVED IN THIS CASE HAS PASSED TESTING WITH NO FAULTS FOUND. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Description of Event or Problem · 1

THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE METER DOES NOT TURN OFF AND THE BUTTONS ARE UNRESPONSIVE. THESE ISSUES WERE NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 3032320

Patients

Seq Age Sex Outcome Treatment
1