FDA Adverse Event Malfunction Summary report: N

SYRINGE 20ML LL S/C 48

MDR report key: 19433988 · Received May 31, 2024

Report

Report Number
1911916-2024-00393
Event Type
Malfunction
Date Received
May 31, 2024
Date of Event
May 28, 2024
Report Date
June 5, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903028307
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

PR 10265349 FOLLOW UP. IT WAS REPORTED THE PRODUCT HAS LITTLE TO NO PRODUCT PRINT ON THEM. TO AID IN THE INVESTIGATION, ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTO SHOWS A PACKAGING BLISTER TOP WEB WITH MISSING PRINT. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR IF THE PACKAGING TOP WEB PRINTER HEADS BECAME CLOGGED INDUCING THE SYMPTOM REPORTED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302830, LOT 4080764. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. VERIFICATION OF THE PACKAGING PRINTING PROCESS WAS PERFORMED. THE PRINTING AND FLOW OF PRODUCT WAS GOOD. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.

Description of Event or Problem · 0

MATERIAL # 302830. LOT # 4080764. IT WAS REPORTED BY CUSTOMER THAT THE 1ST TWO LAYERS OF PRODUCT HAVE LITTLE TO NO PRODUCT PRINT ON THEM. VERBATIM: UPON OPENING A SEALED CASE OF 4 BOXES OF # 302830 ¿ 20ML LL SYRINGE, IT IS NOTED THAT THE 1ST TWO LAYERS OF PRODUCT HAVE LITTLE TO NO PRODUCT PRINT ON THEM. SEE PICS BELOW. LOT # 4080764. WE WILL USE THEM BUT WANTED TO ALERT BD. WE DO HAVE OTHER LOOSE BOXES OF SAME LOT # ON SHELF ALREADY, AND THEY ARE ALL FINE. JUST REPORTING IN CASE OTHER CUSTOMERS REPORT SAME. ADDITIONAL INFORMATION: THE MISS-PRINTED SYRINGE PACKAGING WAS DISCOVERED BY MATERIALS MANAGEMENT STAFF YESTERDAY, 05/28/2024 AND ITEM DIDN¿T MAKE IT TO THE NURSING FLOOR. NO HARM TO PATIENT.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED THE MISS-PRINTED SYRINGE PACKAGING WAS DISCOVERED BY MATERIALS MANAGEMENT STAFF YESTERDAY, 05/28/2024 AND ITEM DIDN¿T MAKE IT TO THE NURSING FLOOR. NO HARM TO PATIENT MATERIAL # 302830. LOT # 4080764. IT WAS REPORTED BY CUSTOMER THAT THE 1ST TWO LAYERS OF PRODUCT HAVE LITTLE TO NO PRODUCT PRINT ON THEM. VERBATIM: UPON OPENING A SEALED CASE OF 4 BOXES OF # 302830 ¿ 20ML LL SYRINGE, IT IS NOTED THAT THE 1ST TWO LAYERS OF PRODUCT HAVE LITTLE TO NO PRODUCT PRINT ON THEM. SEE PICS BELOW. LOT # 4080764. WE WILL USE THEM BUT WANTED TO ALERT BD. WE DO HAVE OTHER LOOSE BOXES OF SAME LOT # ON SHELF ALREADY, AND THEY ARE ALL FINE. JUST REPORTING IN CASE OTHER CUSTOMERS REPORT SAME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299239 SYRINGE 20ML LL S/C 48 SYRINGE, PISTON FMF BECTON DICKINSON 4080764 30382903028307

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown