UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 1943349
·
Received January 1, 2011
Report
- Report Number
- 2050012-2010-01770
- Event Type
- Malfunction
- Date Received
- January 1, 2011
- Date of Event
- December 3, 2010
- Report Date
- December 3, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
QC SHOWED HIGH FLIERS. ON (B)(4) 2010, A BCI FIELD SERVICE ENGINEER (FSE) REPLACED SODIUM (NA) ELECTRODES, SAMPLES PROBE, WASH COLLAR, AND CARBON BRIDGE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUSLY ELEVATED SODIUM (NA) RESULTS INTERMITTENTLY GENERATED BY UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. WHEN DELTA CHECK FAILED, THE SAMPLES WERE REPEATED ON AN ALTERNATE INSTRUMENT. REPEAT TESTING PRODUCED LOWER RESULTS WHICH WERE REPORTED. THERE WAS NO EFFECT TO PATIENTS.