FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 1943349 · Received January 1, 2011

Report

Report Number
2050012-2010-01770
Event Type
Malfunction
Date Received
January 1, 2011
Date of Event
December 3, 2010
Report Date
December 3, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC SHOWED HIGH FLIERS. ON (B)(4) 2010, A BCI FIELD SERVICE ENGINEER (FSE) REPLACED SODIUM (NA) ELECTRODES, SAMPLES PROBE, WASH COLLAR, AND CARBON BRIDGE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUSLY ELEVATED SODIUM (NA) RESULTS INTERMITTENTLY GENERATED BY UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. WHEN DELTA CHECK FAILED, THE SAMPLES WERE REPEATED ON AN ALTERNATE INSTRUMENT. REPEAT TESTING PRODUCED LOWER RESULTS WHICH WERE REPORTED. THERE WAS NO EFFECT TO PATIENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMSTRY ANALYZER JJE BECKMAN COULTER, INC. DXC 600 PRO NA

Patients

Seq Age Sex Outcome Treatment
1