FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 1943326 · Received December 31, 2010

Report

Report Number
2122870-2010-00999
Event Type
Malfunction
Date Received
December 31, 2010
Date of Event
December 1, 2010
Report Date
December 3, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
MMI
PMA / PMN Number
K060256
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS COLLECTED IN A 12 X 75 MM BD LITHIUM HEPARIN PLASMA TUBE WITH A GEL SEPARATOR. THE SAMPLE WAS ANALYZED FROM THE PRIMARY TUBE AND WAS A FULL DRAW WITH NO VISUAL SIGN OF FIBRIN. QC IS PERFORMED EVERY 12 HOURS, AND WAS WITHIN THE ESTABLISHED RANGES. A ROUTINE SYSTEM CHECK WAS PERFORMED ON (B)(6) 2010 AND THE RESULTS MET THE SPECIFICATIONS. A BCI FIELD SERVICE ENGINEER (FSE) WAS ON SITE ON (B)(6) 2010. THE FSE DISCOVERED BUBBLES IN THE WASH PUMP, AND REPLACED SOME HARDWARE COMPONENTS INCLUDING THE WASH PUMP SEALS. THE FSE PRIMED THE SYSTEM AND PERFORMED A HIGH SENSITIVITY SYSTEM CHECK. NO ISSUE WAS NOTED. THE FSE THEN NOTICED AN AIR SLUG FORMING IN THE WASH PUMP, AND REPLACED THE WASH VALVE ROTOR AND THE PRECISION PUMP STATOR. THE FSE PERFORMED A ROUTINE SYSTEM CHECK AND A PRECISION TEST. ALL TESTING MET THE SPECIFICATIONS. ALTHOUGH HARDWARE ISSUES WERE ADDRESSED, A DEFINITIVE ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUSLY ELEVATED TROPONIN (ACCUTNI) RESULTS ABOVE THE AMI CUTOFF GENERATED BY UNICEL DXC 600I ACCESS 2 IMMUNOASSAY SYSTEM FOR ONE PATIENT. SUBSEQUENT TESTING PRODUCED A LOWER RESULT WITHIN THE RISK STRATIFICATION RANGE. THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR OBSERVATION DUE TO THE ERRONEOUS RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600I ACCESS® 2 IMMUNOASSAY SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER MMI BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR Other