RENAL - DISPOSABLE
Report
- Report Number
- 1423500-2010-07483
- Date Received
- December 31, 2010
- Date of Event
- December 8, 2010
- Report Date
- December 8, 2010
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR THE SYSTEM ERROR (SE) 2240 WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. THE ROOT CAUSE WAS NOT IDENTIFIED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR, BECAUSE THE PRODUCT IS UNKNOWN AT THIS TIME. PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED SHOULD ANY SIGNIFICANT SUPPLEMENTAL INFORMATION BECOME AVAILABLE.
A CUSTOMER CONTACTED BAXTER'S (B)(4) REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DWELL 3 OF 4. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM TO THE HOME PATIENT (HP), ASSISTED WITH TROUBLESHOOTING THE ALARM, ADVISED TO USE MANUAL BAG AND NOTIFY HIS NURSE. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE HP. THERE WAS PATIENT INVOLVEMENT BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RENAL - DISPOSABLE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HC CYCLER |