FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 CXE VOLUMETRICINFUSION PUMP

MDR report key: 1943124 · Received December 30, 2010

Report

Report Number
6000001-2010-06509
Event Type
Malfunction
Date Received
December 30, 2010
Date of Event
December 7, 2010
Report Date
December 7, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Removal / Correction Number
600000112/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CONDITION OF BATTERY ISSUE WAS CONFIRMED THROUGH THE EVENT HISTORY. THE CONDITION IS DUE TO DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4)

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). (B)(4)

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER (B)(4) PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED ALARM WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.12.00 CATEGORIZED AS UNREMEDIATED.