8MM MCS TIP COVER ACCESSORY
Report
- Report Number
- 2955842-2010-00561
- Event Type
- Malfunction
- Date Received
- December 30, 2010
- Date of Event
- November 18, 2010
- Report Date
- December 1, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD
- Reporter Occupation
- RISK MANAGER
Narratives
THE TIP COVER ACCESSORY WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, THE TIP COVER WAS FOUND TO HAVE A MECHANICAL TEAR APPROXIMATELY .080 LONG AT THE DISTAL TIP OF THE SILICONE. THE TEAR IS IN THE AXIAL DIRECTION. THE TIP COVER ACCESSORY DOES NOT EXHIBIT ANY DAMAGE RELATED TO ARCING. NO OTHER DAMAGE WAS FOUND. ON (B)(6), 2010 (B)(6), (B)(6) RISK MANAGER, STATED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, ARCING FROM THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WITH THE MCS TIP COVER INSTALLED WAS OBSERVED. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
ON (B)(4), 2010, MEDWATCH UF / IMPORTER REPORT (B)(4) WAS RECEIVED: THE BOVIE CAUTERY UNIT COVER HAD A DEFECT IN IT. WHEN THE CAUTERY WAS ACTIVATED IT ARCED THROUGH THE DEFECT CAUSING A LACERATION OF THE PATIENT'S ILEAC VEIN. THE VEIN WAS REPAIRED. THE COVER DEVICE WAS INSPECTED PRIOR TO USE AND NO TEARS OR PUNCTURES WERE NOTED, HOWEVER, TWO ADDITIONAL EVENTS THAT DID NOT RESULT IN PATIENT HARM OCCURRED THE DAY BEFORE THIS EVENT WITH THE SAME DEVICE. ONE LOOKED LIKE A SMALL TEAR OR PUNCTURE, AND THE OTHER LOOKED LIKE A SMALL SLIT IN THE MATERIAL. THE PACKAGING WAS NOT RETAINED HOWEVER THE LOT FROM THE STORAGE AREA FOR THE SAME PRODUCT WAS C10222. MEDWATCH MFR REPORTS 2955842-2010-00562 AND 2955842-2010-00563 WERE ALSO SENT TO THE FDA REGARDING THESE EVENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 8MM MCS TIP COVER ACCESSORY | ELECTROSURGICAL INSTRUMENT ACCESSORY | NAY | INTUITIVE SURGICAL,INC. | 400180-10 | C10222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYS, INSTRUMENTS, ACCESSORIES, & ESU |