FDA Adverse Event Malfunction Summary report: N

DEXTRUS 4136

MDR report key: 1942948 · Received December 23, 2010

Report

Report Number
1028232-2010-02817
Event Type
Malfunction
Date Received
December 23, 2010
Date of Event
November 4, 2010
Report Date
December 16, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE, BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE LONGEVITY WAS NOT AS LONG AS EXPECTED. THE RIGHT ATRIAL LEAD THRESHOLDS ARE VERY HIGH AND IT IS BELIEVED THAT IS THE CAUSE OF PREMATURE DEPLETION. THE DEVICE WILL BE CHANGED OUT AND THE ATRIAL LEAD MIGHT BE REVISED AT THE SAME TIME. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED. AS OF THIS DATE, THE LEAD HAS NOT BEEN RETURNED. IF THIS LEAD SHOULD BE RETURNED TO OUR COMPANY, IT WILL BE ANALYZED AND THIS INVESTIGATION WILL BE REOPENED AND UPDATED AS NECESSARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4136 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 UNK Other