LIFEVEST WCD 3100 SYSTEM
Report
- Report Number
- 3002158293-2010-01330
- Event Type
- Malfunction
- Date Received
- December 23, 2010
- Date of Event
- November 13, 2010
- Report Date
- December 13, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BENT PINS IN CONNECTOR) HAS BEEN CONFIRMED. UPON INSPECTION, THE ELECTRODE BELT WAS FOUND TO HAVE THE CONNECTOR PINS BENT IN A SWIRL PATTERN. THE ELECTRODE BELT CONNECTOR PINS DID NOT SUCCESSFULLY MATE WITH THE CONNECTOR. THE ROOT CAUSE OF THE BENT PINS CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY DUE TO THE CONNECTOR BEING FORCED INTO THE MONITOR WHILE THE PINS WERE MISALIGNED. NO ADVERSE EVENT RESULTED FROM THE BENT CONNECTOR PINS. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A PSR ASSISTING A (B)(6) FEMALE PATIENT CALLED IN TO ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT THAT THERE WERE BENT PINS IN THE PATIENT'S ELECTRODE BELT CONNECTOR, CAUSING THE MONITOR NOT TO RECOGNIZE THE BELT CONNECTION. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |