FDA Adverse Event Malfunction Summary report: N

S8 AUTOSET VANTAGE - AMERICAS

MDR report key: 1942904 · Received December 23, 2010

Report

Report Number
3004604967-2010-00047
Event Type
Malfunction
Date Received
December 23, 2010
Date of Event
November 1, 2010
Report Date
December 23, 2010
Manufacturer
RESMED LTD
Product Code
BZD
PMA / PMN Number
K041209
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FURTHER CONTACT WITH THE INITIAL REPORTER REVEALED THAT THE PT DID NOT SEEK MEDICAL ATTENTION. THE REPORTED COMPLAINT OF BURN WAS ACTUALLY A REDNESS AND TENDERNESS AROUND THE MOUTH AREA. THE ENGINEERING EVALUATION OF THE RETURNED UNIT IDENTIFIED THE ROOT CAUSE OF THE FAILURE AS THE AC APPLIANCE INLET CONNECTOR SOLDER JOINT. THE RESULT OF THIS FAILURE WAS A BRIEF EXTERNAL FLAME THAT SELF-ARRESTED AND DID NOT REQUIRE EXTERNAL INTERVENTION. THE INCIDENCE AND SEVERITY OF THIS FAILURE TYPE ARE SUBJECTED TO ON-GOING MONITORING.

Description of Event or Problem · 1

IT WAS REPORTED THAT A RESMED CPAP UNIT CAUGHT ON FIRE AT THE POWER CORD AREA. THE INITIAL REPORT ALLEGED THAT THE PT HAD A BURN ON HIS FACE AND COUGHED UP BLACK MATERIAL, HOWEVER, HE DID NOT REQUIRE ANY MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 S8 AUTOSET VANTAGE - AMERICAS BZD RESMED LTD 33112

Patients

Seq Age Sex Outcome Treatment
1