S8 AUTOSET VANTAGE - AMERICAS
Report
- Report Number
- 3004604967-2010-00047
- Event Type
- Malfunction
- Date Received
- December 23, 2010
- Date of Event
- November 1, 2010
- Report Date
- December 23, 2010
- Manufacturer
- RESMED LTD
- Product Code
- BZD
- PMA / PMN Number
- K041209
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
FURTHER CONTACT WITH THE INITIAL REPORTER REVEALED THAT THE PT DID NOT SEEK MEDICAL ATTENTION. THE REPORTED COMPLAINT OF BURN WAS ACTUALLY A REDNESS AND TENDERNESS AROUND THE MOUTH AREA. THE ENGINEERING EVALUATION OF THE RETURNED UNIT IDENTIFIED THE ROOT CAUSE OF THE FAILURE AS THE AC APPLIANCE INLET CONNECTOR SOLDER JOINT. THE RESULT OF THIS FAILURE WAS A BRIEF EXTERNAL FLAME THAT SELF-ARRESTED AND DID NOT REQUIRE EXTERNAL INTERVENTION. THE INCIDENCE AND SEVERITY OF THIS FAILURE TYPE ARE SUBJECTED TO ON-GOING MONITORING.
IT WAS REPORTED THAT A RESMED CPAP UNIT CAUGHT ON FIRE AT THE POWER CORD AREA. THE INITIAL REPORT ALLEGED THAT THE PT HAD A BURN ON HIS FACE AND COUGHED UP BLACK MATERIAL, HOWEVER, HE DID NOT REQUIRE ANY MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | S8 AUTOSET VANTAGE - AMERICAS | BZD | RESMED LTD | 33112 |