FDA Adverse Event Malfunction Summary report: N

CONCERTO, BASIC & SHOWER TROLLEY

MDR report key: 1942815 · Received December 20, 2010

Report

Report Number
9611530-2010-00127
Event Type
Malfunction
Date Received
December 20, 2010
Date of Event
December 16, 2010
Report Date
December 16, 2010
Manufacturer
ARJO HOSPITAL EQUIPMENT AB
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WE ARE REPORTING ACCORDING TO EXEMPTION # (B)(4). INCIDENTS INVOLVING MEDICAL DEVICES MANUFACTURED BY ARJO HOSPITAL EQUIPMENT AB IN (B)(4) WILL BE REPORTED BY US, THE LEGAL MANUFACTURER, ARJO HOSPITAL EQUIPMENT AB IN (B)(4) ON BEHALF OF OUR SALES AND DISTRIBUTION COMPANY IN (B)(4). ADDITIONAL INFO WILL BE PROVIDED UPON CONCLUSION OF THE MANUFACTURER'S INVESTIGATION.

Description of Event or Problem · 1

AS STATED BY THE CUSTOMER ON (B)(6) 2010: SERIAL # OF STRETCHER IS (B)(4). SOME SCREWS WERE LOOSEN OUT AND THE STRETCHER DETACHED FROM THE CHASSIS SUDDENLY WITH PT. LUCKILY THE NURSE HELD THE STRETCHER AND THE PT DIDN'T FALL TO THE FLOOR. ANOTHER SIDE OF HOLDER HAS 1 SCREW LOOSEN AND LOST. THIS BED WAS USED ONLY 2 YEARS AND THE CONCERTO OVERALL CONDITION WAS STILL GOOD. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO, BASIC & SHOWER TROLLEY LIFT, PATIENT, NON-AC-POWERED FSA ARJO HOSPITAL EQUIPMENT AB BAB1000-01

Patients

Seq Age Sex Outcome Treatment
1