FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1942797
·
Received December 20, 2010
Report
- Report Number
- 2531779-2010-02983
- Event Type
- Malfunction
- Date Received
- December 20, 2010
- Report Date
- November 21, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, THE CONTRAST KEYPAD BUTTON UNRESPONSIVE TO USER INPUT WHEN PRESSED, THE UP AND DOWN KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE WHEN PRESSED, THE CONTRAST BUTTON ALSO WAS STICKING WHEN PRESSED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE CONTRAST, UP AND DOWN ARROW KEYPAD BUTTONS INTERMITTENTLY UNRESPONSIVE WHEN PRESSED. THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |