FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1942797 · Received December 20, 2010

Report

Report Number
2531779-2010-02983
Event Type
Malfunction
Date Received
December 20, 2010
Report Date
November 21, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, THE CONTRAST KEYPAD BUTTON UNRESPONSIVE TO USER INPUT WHEN PRESSED, THE UP AND DOWN KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE WHEN PRESSED, THE CONTRAST BUTTON ALSO WAS STICKING WHEN PRESSED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE CONTRAST, UP AND DOWN ARROW KEYPAD BUTTONS INTERMITTENTLY UNRESPONSIVE WHEN PRESSED. THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020

Patients

Seq Age Sex Outcome Treatment
1