ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2010-02978
- Event Type
- Malfunction
- Date Received
- December 20, 2010
- Report Date
- November 20, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- UNKNOWN
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, THE UP, DOWN AND OK KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO USER INPUTS DURING TESTING, THE CONTRAST KEYPAD BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, IT WAS OBSERVED THAT THE UP AND OK KEY CONTACTS WERE MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED, THE UP, DOWN, AND OK KEY CONTACTS BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL KEY CONTACTS.
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP, DOWN, AND OK KEYPAD BUTTONS UNRESPONSIVE WHEN PRESSED. IT WAS REPORTED THAT THE BUTTONS FEEL FLAT, HOWEVER, THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |