FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1942790 · Received December 20, 2010

Report

Report Number
2531779-2010-02977
Event Type
Malfunction
Date Received
December 20, 2010
Date of Event
September 27, 2010
Report Date
November 22, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010/003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE. THE PUMP WAS PRIMED SUCCESSFULLY AND EXERCISED WITH NO ALARMS OCCURRING.

Description of Event or Problem · 1

EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR