FDA Adverse Event Injury Summary report: N

LOGIC CR FEMORAL CEM, LEFT SZ 2

MDR report key: 19427847 · Received May 30, 2024

Report

Report Number
1038671-2024-01605
Event Type
Injury
Date Received
May 30, 2024
Date of Event
April 18, 2024
Report Date
September 19, 2024
Manufacturer
EXACTECH, INC.
Product Code
JWH
PMA / PMN Number
K111400
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION H10: (H3) PENDING EVALUATION, (D10) CONCOMITANT DEVICE(S): 4632315 - 02-012-45-2020 - LGC TIBIAL FIT TRAY CEM SZ 2F / 2T. 6896919 - 200-02-35 - THREE PEG PATELLA 35MM. 6176661 - 201-78-15 - HOLDING PIN MINI SHARP POINT 4 PK. 6837191 - 201-78-81 - 3 TROCAR, MOD. HEX 2PK. S337427 - 521-78-23 - THREADED PIN SIZE 2.3 COLLARED. S337437 - 521-78-23 - THREADED PIN SIZE 2.3 COLLARED. 4579701 - 521-78-32 - THREADED PIN SIZE 3.0 COLLARLESS. 9000221286 - A10012 - GPS IMPLANT KIT V2. 6108425 - 02-012-51-2009 - LOGIC TIB INSERT IMPL CRC, SZ 2, 9MM.

Additional Manufacturer Narrative · 0

(H3) THE REVISION REPORTED WAS LIKELY THE RESULT OF PRESUMED POLYETHYLENE WEAR AND AN INSUFFICIENT BOND BETWEEN THE FEMORAL COMPONENT AND THE BONE, WHICH MAY HAVE LED TO ASEPTIC (NON-INFECTED) FEMORAL LOOSENING. HOWEVER, THE REPORTED FEMORAL LOOSENING COULD NOT BE CONFIRMED AS THE DEVICES WERE NOT RETURNED FOR EVALUATION. *THE FOLLOWING SECTIONS HAVE CORRECTED INFORMATION: (H6) HEALTH EFFECT - CLINICAL CODE: 2377, OSTEOLYSIS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS 60 Y/O FEMALE PATIENT'S LEFT KNEE WAS REVISED APPROXIMATELY 2 YEARS POST OP. FEMUR AND TIBIA WERE REVISED DUE TO OSTEO-LYSIS AND FEMORAL LOOSENING. SURGEON REMOVED ALL EXACTECH DEVICES AND PUT ANOTHER COMPANIES PRODUCT IN. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PRODUCT NOT RETURNING - DISCARDED AT THE HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1225050 LOGIC CR FEMORAL CEM, LEFT SZ 2 PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. UNK

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female Required Intervention SEE H10.