VITALHOLD STD BORE EXTSET W/VLCATHETER SECUREMENT (CONVENIENCE KIT)
Report
- Report Number
- 6000001-2010-06420
- Event Type
- Malfunction
- Date Received
- December 30, 2010
- Date of Event
- March 1, 2010
- Report Date
- March 30, 2010
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- OJA
- PMA / PMN Number
- K081289
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE CUSTOMER SENT IN FIVE USED SAMPLES FOR EVALUATION. ALL SAMPLES WERE ATTACHED TO AN IN-HOUSE PRIMARY SET WHICH WAS SPIKED INTO A 1000ML SOLUTION BAG CONTAINING REVERSE OSMOSIS WATER. SAMPLES 1-4 PRIMED NORMALLY WITH NORMAL FLOW NOTED. HOWEVER, SAMPLE 5 WOULD NOT PRIME. A VISUAL INSPECTION UNDER A MICROSCOPE OF THE GLAND ON SAMPLE 5 REVEALED A RE-KNIT CONDITION OF THE CENTER SLIT. BASED ON THE EVALUATION RESULTS THE CONDITION WAS CONFIRMED. THIS CONDITION WAS CAUSED BY RE-KNIT OF THE CENTER SLIT. THERE IS NO 510K NUMBER ASSOCIATED WITH THIS PRODUCT AS IT IS 510K EXEMPT. HOWEVER, THE INDIVIDUAL COMPONENT FROM THE KIT WITH THE REPORTED DEFECT HAS A 510K NUMBER. THEREFORE, IN THIS MDR, THE 510K NUMBER UTILIZED IS REFERRING TO THE SPECIFIC COMPONENT OF THE KIT RELATING TO THE REPORTED CONDITION. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE CUSTOMER REPORTED TO BAXTER (B)(4) A VITALHOLD STANDARD BORE EXTENSION SET WITH V-LINK CATHETER SECUREMENT (CONVENIENCE KIT) THAT WOULD NOT ALLOW FLOW THROUGH THE V-LINK AFTER THE SOLUTION BAG WAS SPIKED. THE CUSTOMER WAS USING A CONTINU-FLO 109" INCH PRIMARY SET IN CONJUNCTION WITH THE V-LINK EXTENSION SET. THE CONDITION OCCURRED DURING PRIMING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALHOLD STD BORE EXTSET W/VLCATHETER SECUREMENT (CONVENIENCE KIT) | INTRAVENOUS EXTENSION TUBING SET | OJA | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CONTINU-FLO 109" INCH PRIMARY SET |