FDA Adverse Event Malfunction Summary report: N

SAFE GAIT ACTIVE 360 DYNAMIC MOBILITY TRAINER

MDR report key: 19426441 · Received May 30, 2024

Report

Report Number
3030997791-2024-00001
Event Type
Malfunction
Date Received
May 30, 2024
Date of Event
May 2, 2024
Report Date
May 30, 2024
Manufacturer
DIH TECHNOLOGIES INC.
Product Code
BXB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IN THIS LOT WAS 1 OF 4 MANUFACTURED BY A SUPPLIER NO LONGER USED BY DIH. THE ASSEMBLY OF THE SPREADER BAR HAS MOVED TO OUR OWN MANUFACTURING FACILITY WHERE QA PROCEDURES ARE IN PLACE TO DETECT DEFECTIVE SELF-LOCKING NUTS. SINCE THE SUSPECT LOT OF DEVICES HAVE BEEN INSPECTED, WE HAVE DETERMINED THAT IT IS IMPROBABLE FOR THIS EVENT TO REOCCUR.

Description of Event or Problem · 0

DIH TECHNOLOGIES, INC. RECEIVED A REPORT STATING THE CENTER BOLT ON YOKE OF THE SPREADER BAR BECAME DETACHED DURING A PATIENT SESSION. THE CAREGIVER WAS ABLE TO STABILIZE THE PATIENT DURING THE EVENT. NO ADVERSE EVENT OCCURRED. THE SPREADER BAR WAS REPLACED ON THIS DEVICE. THE SPREADER BAR ASSEMBLY WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE PRIMARY ROOT CAUSE WAS DETERMINED THAT THE SELF-LOCKING NUT DESIGNED TO PREVENT THE SPREADER BAR FROM DETACHING WAS DEFECTIVE. THE DEFECTIVE NUT WAS FOUND TO HAVE GAPS THAT WERE OUT OF SPECIFICATION. THESE GAPS DID NOT ALLOW THE PROPER ENGAGEMENT WITH THE CENTER BOLT. IN NORMAL OPERATION, THE SPREADER BAR CAN SPIN 360 DEGREES ON A BEARING THAT IS INSTALLED ABOVE THE SELF-LOCKING NUT. SINCE THE SELF-LOCKING NUT WAS NOT ENGAGING WITH THE BOLT SUFFICIENTLY, EXPECTED FORCES APPLIED FROM THE PATIENT'S WEIGHT ALLOWED THE NUT TO BACK-OFF AS THE SPREADER BAR ROTATED. SPREADER BAR ASSEMBLIES FROM THIS LOT WERE INSPECTED FOR DEFECTIVE NUTS AND NO ADDITIONAL DEFECTS WERE FOUND. THE ANNUAL MAINTENANCE RECORDS OF ALL OTHER FIELD UNITS DO NOT SHOW ANY DEFECTIVE NUTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1100180 SAFE GAIT ACTIVE 360 DYNAMIC MOBILITY TRAINER POWERED EXERCISE EQUIPMENT BXB DIH TECHNOLOGIES INC.

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown