FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM

MDR report key: 1942640 · Received December 30, 2010

Report

Report Number
2050012-2010-01762
Event Type
Malfunction
Date Received
December 30, 2010
Date of Event
November 10, 2010
Report Date
December 1, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER STATED THAT QC IS LOW WHEN RERUN AFTER LOW SODIUM (NA) RESULTS ARE NOTICED. IF THE ANION GAP IS LOW, THEN THE SAMPLES ARE RERUN. THE ANION GAPS WERE 4-5 UNITS LOWER ON THIS INSTRUMENT FOR THE SAMPLES. A BCI FIELD SERVICE ENGINEER NOTED ON (B)(4) 2010 THAT THE CUSTOMER HAD CHANGED THE NA MEASURING ELECTRODE AND IT APPEARED TO HAVE RESOLVED THE ISSUE. THE FSE VISITED THE SITE ON (B)(4) 2010, AND PERFORMED VERIFICATION TEST. ON (B)(4) 2010, THE FSE DECONTAMINATED THE INSTRUMENT, REPLACED NA AND CARBONE DIOXIDE (CO2) ELECTRODES, SAMPLE PROBE, AND WASH COLLAR.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUSLY LOW SODIUM (NA) RESULTS INTERMITTENTLY GENERATED BY UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM BETWEEN (B)(6) 2010 AND (B)(6) 2010. THE RESULTS WERE REPORTED OUT OF THE LABORATORY. RERUNS WERE TRIGGERED BY LOW ANION GAPS. THE REPEAT RESULTS AGREED WITH ANOTHER INSTRUMENT IN THE LAB. BOTH INITIAL AND REPEAT RESULTS WERE ALL WITHIN THE NORMAL REFERENCE RANGE. PATIENT TREATMENT WAS NOT INITIATED OR WITHHELD UPON THE ERRONEOUS RESULTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM CLINICAL CHEMSTRY ANALYZER JJE BECKMAN COULTER, INC. DXC 800 PRO NA

Patients

Seq Age Sex Outcome Treatment
1