FDA Adverse Event Malfunction Summary report: N

HARMONIC CURVED SHEARS INSERT

MDR report key: 1942604 · Received December 30, 2010

Report

Report Number
2955842-2010-00565
Event Type
Malfunction
Date Received
December 30, 2010
Date of Event
November 10, 2010
Report Date
December 1, 2010
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K011281
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

THE INSERT ACCESSORY WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND THAT THE CLAMP ARM DETACHED FROM THE DISTAL END. ONE SIDE OF THE SHAFT FEATURE THAT MATES WITH CLAMP ARM PINS IS BENT BACKWARDS. THIS DAMAGE IS INDICATIVE OF CLAMP ARM HYPEREXTENSION. WHILE THE EVIDENCE IS INCONCLUSIVE, ENGINEERING CONCLUDED THAT DAMAGE TO THE INSERT ACCESSORY IS MOST LIKELY DUE TO MISHANDLING/MISUSE. ON (B)(6), 2010 (B)(6), (B)(6) HOSPITAL RISK MANAGER STATED THAT DURING A DA VINCI S COLOSTOMY CLOSURE PROCEDURE, A PIECE OF THE HARMONIC CURVED SHEAR (HCS) INSERT BROKE OFF AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED. SHE ALSO STATED THAT A PIECE FROM A REPLACEMENT INSERT ACCESSORY ALSO USED DURING THE SURGICAL PROCEDURE BROKE OFF AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY. MED WATCH REPORT 2955842-2010-00564 WAS ALSO SUBMITTED REGARDING THIS EVENT.

Description of Event or Problem · 1

ON (B)(4), 2010, ISI RECEIVED UF/IMPORTER REPORT (B)(4) WITH THE FOLLOWING EVENT DESCRIPTION: END PIECE BROKE OFF AND WAS RETRIEVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC CURVED SHEARS INSERT ENDOSCOPIC ACCESSORY NAY INTUITIVE SURGICAL,INC. 400169-07 M10100802

Patients

Seq Age Sex Outcome Treatment
1 80 YR DA VINCI S SURG SYSTEM, HCS INSTRUMENT, ACCSS, ES