HARMONIC CURVED SHEARS INSERT
Report
- Report Number
- 2955842-2010-00565
- Event Type
- Malfunction
- Date Received
- December 30, 2010
- Date of Event
- November 10, 2010
- Report Date
- December 1, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K011281
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- RISK MANAGER
Narratives
THE INSERT ACCESSORY WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND THAT THE CLAMP ARM DETACHED FROM THE DISTAL END. ONE SIDE OF THE SHAFT FEATURE THAT MATES WITH CLAMP ARM PINS IS BENT BACKWARDS. THIS DAMAGE IS INDICATIVE OF CLAMP ARM HYPEREXTENSION. WHILE THE EVIDENCE IS INCONCLUSIVE, ENGINEERING CONCLUDED THAT DAMAGE TO THE INSERT ACCESSORY IS MOST LIKELY DUE TO MISHANDLING/MISUSE. ON (B)(6), 2010 (B)(6), (B)(6) HOSPITAL RISK MANAGER STATED THAT DURING A DA VINCI S COLOSTOMY CLOSURE PROCEDURE, A PIECE OF THE HARMONIC CURVED SHEAR (HCS) INSERT BROKE OFF AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED. SHE ALSO STATED THAT A PIECE FROM A REPLACEMENT INSERT ACCESSORY ALSO USED DURING THE SURGICAL PROCEDURE BROKE OFF AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY. MED WATCH REPORT 2955842-2010-00564 WAS ALSO SUBMITTED REGARDING THIS EVENT.
ON (B)(4), 2010, ISI RECEIVED UF/IMPORTER REPORT (B)(4) WITH THE FOLLOWING EVENT DESCRIPTION: END PIECE BROKE OFF AND WAS RETRIEVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC CURVED SHEARS INSERT | ENDOSCOPIC ACCESSORY | NAY | INTUITIVE SURGICAL,INC. | 400169-07 | M10100802 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | DA VINCI S SURG SYSTEM, HCS INSTRUMENT, ACCSS, ES |