FDA Adverse Event Malfunction Summary report: N

HD 3CMOS AUTOCLAVABLE CAMERA HEAD

MDR report key: 19425968 · Received May 30, 2024

Report

Report Number
3002808148-2024-33519
Event Type
Malfunction
Date Received
May 30, 2024
Date of Event
May 14, 2024
Report Date
June 27, 2024
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
PMA / PMN Number
K200542
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. NEW FAILURE PHENOMENA IDENTIFIED DURING THE INSPECTION: WATERTIGHT INSPECTION NG A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: WATERTIGHT INSPECTION NG A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: COMPONENT FAILURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL THE EVALUATION OF THE EVENT IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CABLE ON THE CAMERA HEAD HAD A LOOSE FIT. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CABLE ON THE SUBJECT DEVICE HAD A LOOSE FIT. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1423459 HD 3CMOS AUTOCLAVABLE CAMERA HEAD AUTOCLAVABLE CAMERA HEAD FET SHIRAKAWA OLYMPUS CO., LTD. CH-S200-XZ-EA

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown