ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2010-03088
- Event Type
- Malfunction
- Date Received
- December 23, 2010
- Report Date
- November 26, 2010
- Manufacturer
- ANIMAS CORP
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT NO LIFTING OR PEELING. THE "UP/DOWN/OK/CONTRAST" KEYPAD BUTTONS ARE INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING. THE KEYPAD WAS REMOVED FOR FURTHER INVESTIGATION. INSPECTION, THE "UP/DOWN/OK/CONTRAST" KEY CONTACTS ARE MISALIGNED, THE "CONTRAST" KEY CONTACT IS INVERTED, AND THE "UP/DOWN/OK" KEY CONTACTS BECOMES INVERTED WHEN PRESSED AND DO NOT ALWAYS CLICK AND SPRING BACK. THERE WAS EVIDENCE KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. REPLACEMENT PRODUCTS WERE SENT TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |