NV GXL LINR, NTRL, 36MM ID, GROUP 3 CUPS
Report
- Report Number
- 1038671-2024-01578
- Event Type
- Injury
- Date Received
- May 30, 2024
- Date of Event
- November 6, 2023
- Report Date
- January 31, 2025
- Manufacturer
- EXACTECH, INC.
- Product Code
- JDI
- PMA / PMN Number
- K070479
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- 003
Narratives
D10: CONCOMITANT DEVICES: 2855870 180-11-56 - NV CROWN CUP CLSR HL W/HA 56 GROUP 3. 2886583 142-36-00 - COCR FEM HEAD 36MM +0 OFFSET 12/14. 2921426 164-11-12 - NOVATION ELEMENT RO S/O SZ 12. 2937929 120-65-30 - BONE SCREW 6.5MM DIA X 30MM LONG. THE PRODUCT ASSOCIATED WITH THE REPORTED EVENT IS WITHIN THE SCOPE OF RECALL Z-1729-2022; HOWEVER, THERE IS INSUFFICIENT INFORMATION TO EVALUATE WHETHER THE SUBJECT ISSUE OF THE RECALL WAS THE CAUSE OR CONTRIBUTOR TO THE REPORTED EVENT. THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO MEDICAL OR OTHER RECORDS CONTAINING TREATMENT INFORMATION OR PATIENT INFORMATION HAVE BEEN RECEIVED; THEREFORE, THE REPORTED EVENT CANNOT BE CONFIRMED, NOR CAN THE CIRCUMSTANCES OR POTENTIAL CAUSES OR CONTRIBUTORS TO THE ALLEGED EVENT BE EVALUATED. SHOULD ADDITIONAL MATERIAL INFORMATION BECOME AVAILABLE THAT PERMITS MORE ANALYSIS OR CONCLUSIONS, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE FOLLOWING SECTIONS WERE CORRECTED: H6. THE MOST LIKELY UNDERLYING CAUSE FOR THE REVISION REPORTED IS PROSTHESIS WEAR, LOSS OF RANGE OF MOTION, AND ACETABULAR LOOSENING. AND A COMBINATION OF RISK FACTORS SUCH AS USE ERROR, IMPLANT POSITIONING, IMPLANT SIZE SELECTION, AND PATIENT FACTORS (FITNESS FOR SURGERY, BIOMECHANICS, ACTIVITY LEVEL AND LOCAL TISSUE OXIDATION POTENTIAL). HOWEVER, THIS CANNOT BE CONFIRMED FROM THE REPORTED INFORMATION AND THE DEVICES WERE NOT AVAILABLE FOR EVALUATION. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 117 MONTHS AFTER A RIGHT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS ACETABULAR LOOSENING, PAIN, SWELLING, STIFFNESS, LIMITED RANGE OF MOTION, LIMITED MOBILITY, AND PERIACETABULAR OSTEOLYSIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1258725 | NV GXL LINR, NTRL, 36MM ID, GROUP 3 CUPS | PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED | JDI | EXACTECH, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Male | Required Intervention| H | SEE H10. |