FDA Adverse Event Malfunction Summary report: N

REPAIR U2 DRILL

MDR report key: 1942365 · Received December 22, 2010

Report

Report Number
1811755-2010-02095
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
October 26, 2010
Report Date
October 26, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED BY THE MFR AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED AND CORROSION WAS DISCOVERED IN THE MOTOR ASSEMBLY, ROTOR ASSEMBLY, AND BEARINGS. ALL CORRODED COMPONENTS WERE REPLACED AND THE DRILL WAS RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED BY THE USER FACILITY THAT THE DRILL OPERATED WHEN THE TRIGGER WAS NOT ACTIVATED. NO FURTHER INFO IS KNOWN. ATTEMPTS TO OBTAIN ADD'L INFO HAVE BEEN UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REPAIR U2 DRILL DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC. ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK