FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1942340 · Received December 23, 2010

Report

Report Number
2531779-2010-03055
Event Type
Malfunction
Date Received
December 23, 2010
Date of Event
November 22, 2010
Report Date
November 22, 2010
Manufacturer
ANIMAS CORP
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. DURING EVAL THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. EVAL REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.

Description of Event or Problem · 1

DURING EVAL THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. THE FORCE SENSOR PINS WERE FOUND TO BE PARTIALLY DISLODGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 36 YR