FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1942337
·
Received December 22, 2010
Report
- Report Number
- 2531779-2010-03061
- Event Type
- Malfunction
- Date Received
- December 22, 2010
- Report Date
- November 29, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THERE WAS NO DAMAGE TO KEYPAD OBSERVED. THE 'UP/DOWN/CONTRAST' BUTTONS WAS RESPONSIVE AND SPRING BACK WHEN PRESSED. THE 'OK' BUTTON WAS UNRESPONSIVE AND DOES NOT SPRINGS BACK WHEN PRESSED. THE "OK" BUTTON WAS NOTED TO BE INVERTED WHEN KEYPAD WAS REMOVED WHILE THE 'DOWN/CONTRAST/UP' BUTTON CONTACTS ARE NORMAL. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. REPLACEMENT PRODUCTS WERE SENT TO THE PT.