FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1942337 · Received December 22, 2010

Report

Report Number
2531779-2010-03061
Event Type
Malfunction
Date Received
December 22, 2010
Report Date
November 29, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THERE WAS NO DAMAGE TO KEYPAD OBSERVED. THE 'UP/DOWN/CONTRAST' BUTTONS WAS RESPONSIVE AND SPRING BACK WHEN PRESSED. THE 'OK' BUTTON WAS UNRESPONSIVE AND DOES NOT SPRINGS BACK WHEN PRESSED. THE "OK" BUTTON WAS NOTED TO BE INVERTED WHEN KEYPAD WAS REMOVED WHILE THE 'DOWN/CONTRAST/UP' BUTTON CONTACTS ARE NORMAL. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. REPLACEMENT PRODUCTS WERE SENT TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020

Patients

Seq Age Sex Outcome Treatment
1