Additional Manufacturer Narrative · 1
THE HANDPIECE HAS BEEN RECEIVED AT THE MANUFACTURER FOR INVESTIGATION. AN EVALUATION WAS CONDUCTED AND THE COMPLAINT WAS CONFIRMED. ACCORDING TO THE INVESTIGATION DETAILS, THE ROTOR WAS LODGED IN THE MOTOR. THE MOTOR, ROTOR, ALONG WITH OTHER COMPONENTS, WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THERE WAS NO REPORTED PATIENT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.