FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1942275
·
Received December 22, 2010
Report
- Report Number
- 2531779-2010-03067
- Event Type
- Malfunction
- Date Received
- December 22, 2010
- Report Date
- November 29, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE DOWN AND OK BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS, THE KEY CONTACTS WERE INVERTED DURING TESTING AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DOWN ARROW AND OK BUTTONS WERE INTERMITTENTLY ACTIVATING THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |