FDA Adverse Event Malfunction Summary report: N

PUMP MMT-723PNAL PRDGM INS V2.3 CL EN PW

MDR report key: 1942248 · Received December 22, 2010

Report

Report Number
2032227-2010-83534
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
December 9, 2010
Report Date
December 9, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS FOR THE PAST WEEK. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT FAILED THE HIGH PRESSURE TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-723PNAL PRDGM INS V2.3 CL EN PW INFUSION PUMP MDS MEDTRONIC MINIMED MMT-723PNAL

Patients

Seq Age Sex Outcome Treatment
1 14 YR