FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1942184 · Received December 21, 2010

Report

Report Number
2531779-2010-03015
Event Type
Malfunction
Date Received
December 21, 2010
Report Date
November 23, 2010
Manufacturer
ANIMAS CORP
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, ALL BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS, WERE FOUND TO BE MISALIGNED AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DOWN ARROW BUTTON HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1