FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1942180
·
Received December 21, 2010
Report
- Report Number
- 2531779-2010-03003
- Event Type
- Malfunction
- Date Received
- December 21, 2010
- Report Date
- November 22, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO VISUAL DAMAGE TO KEYPAD. 'OK' BUTTON INTERMITTENTLY WAS RESPONSIVE AND 'UP/DOWN/CONTRAST' BUTTONS WERE RESPONSIVE TO USER INPUT. ALL BUTTONS SPRING BACK AS NORMAL WHEN PRESSED. IT WAS NOTED THAT THERE WAS CONTAMINATION AROUND 'OK' BUTTON CONTACT OBSERVED.
Description of Event or Problem · 1
THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|