FDA Adverse Event
Malfunction
Summary report: N
PLUM XL3M W/NURS CAL
MDR report key: 1942137
·
Received November 15, 2010
Report
- Report Number
- 2921482-2010-00892
- Event Type
- Malfunction
- Date Received
- November 15, 2010
- Date of Event
- October 20, 2010
- Report Date
- October 21, 2010
- Manufacturer
- HOSPIRA, INC.
- Product Code
- FRN
- PMA / PMN Number
- K952799
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS REC'D. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL; (B)(4).
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE ON THE LOW SETTING DURING AN ALARM CONDITION. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLUM XL3M W/NURS CAL | 80FRN | FRN | HOSPIRA, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |