AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2010-07406
- Event Type
- Malfunction
- Date Received
- December 30, 2010
- Date of Event
- December 6, 2010
- Report Date
- December 6, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 WAS NOT CONFIRMED DUE TO THE LACK OF A SAMPLE; HOWEVER, BASED ON THE INFORMATION GATHERED DURING BAXTER'S INVESTIGATION, THE CAUSE OF THE SYSTEM ERROR 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT IMPROPERLY DISCONNECTED HIMSELF FROM THE HOMECHOICE AND THEN RECONNECTED. THE LOT INFORMATION WAS UNKNOWN; THEREFORE A BATCH REVIEW WAS NOT PERFORMED. REVIEW OF THE LABELING REVEALS THE HOMECHOICE APD SYSTEMS AT-HOME GUIDE CONTAINS SUFFICIENT LABELING FOR THE USER ERROR IN THIS COMPLAINT. BAXTER HAS RECEIVED SIMILAR REPORTS. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). DURING INVESTIGATION BY BAXTER THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE, THE DEVICE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION.
A CAREGIVER CONTACTED (B)(4) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING DRAIN 3 OF 6. (B)(4) EXPLAINED THAT A LARGE AMOUNT OF AIR HAD ENTERED INTO THE DISPOSABLE SET. THE PATIENT DISCONNECTED HIMSELF AND THEN RECONNECTED. (B)(4) ADVISED THE CAREGIVER THAT THERAPY WAS OVER AND HAD THEM CYCLE POWER TWICE TO CLEAR THE ERROR. (B)(4) THEN ADVISED THE CAREGIVER TO NOTIFY THE PATIENT'S DIALYSIS NURSE. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 92 YR | HOMECHOICE PRO APD CYCLER (B)(4) (REFURBISHED) |