FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 1941920 · Received December 23, 2010

Report

Report Number
1028232-2010-02840
Event Type
Injury
Date Received
December 23, 2010
Date of Event
November 8, 2010
Report Date
December 16, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFO THAT THIS VENTRICULAR LEAD DISLODGED DUE TO NOT ENOUGH SLACK IN THE LEAD. DUE TO TISSUE BEING CLOGGED IN THE HELIX, THE LEAD WAS EXPLANTED AND REPLACED. NO ADVERSE PT EFFECTS WERE REPORTED. THERE IS NO FURTHER INFO AVAILABLE. SHOULD ADDITIONAL INFO BECOME AVAILABLE, THIS EVENT WOULD BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4136 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization