FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 1941908 · Received December 23, 2010

Report

Report Number
2531779-2010-03091
Event Type
Injury
Date Received
December 23, 2010
Date of Event
November 24, 2010
Report Date
November 24, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. A SAMPLE CARTRIDGE WITH THE SAME LOT NUMBER FROM THE PT'S ORDER HISTORY WAS TESTED. EVAL REVEALED NO VISIBLE CARTRIDGE DAMAGE. THERE WERE NO AIR BUBBLES DUPLICATED UPON A FILL TEST AND THERE WERE NO LEAKS OBSERVED WHEN PERFORMING A LEAK TEST. THE ANIMAS REP NOTED THAT THE PT'S TECHNIQUE WHEN FILLING THE CARTRIDGE MAY HAVE CAUSED AIR BUBBLES IN THE CARTRIDGE. THE ANIMAS REP ADVISED THE REPORTER TO MAKE SURE THE CARTRIDGE CONNECTION TO THE LUER WAS TIGHT. THERE WERE NO LEAKS OBSERVED WHEN THE REP TROUBLESHOT WITH THE REPORTER.

Description of Event or Problem · 1

THE PT'S MOTHER REPORTED THAT HER DAUGHTER WENT TO THE EMERGENCY ROOM WITH A BLOOD GLUCOSE LEVEL OF OVER 600 MG/DL AND LARGE KETONES AND NAUSEA. WHEN THE PT ARRIVED AT THE HOSPITAL, SHE NOTICED AIR BUBBLES IN THE INFUSION SET TUBING AND THE CARTRIDGE. SHE ALSO NOTICED THAT INSULIN WAS LEAKING FROM THE LUER CONNECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN CARTRIDGE LZG ANIMAS CORP. IR 1200 / 1250/ 2020/ OTP UNK

Patients

Seq Age Sex Outcome Treatment
1 14 YR Hospitalization| L| R