FDA Adverse Event Injury Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1941779 · Received December 21, 2010

Report

Report Number
2531779-2010-03020
Event Type
Injury
Date Received
December 21, 2010
Date of Event
November 23, 2010
Report Date
November 23, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE THE RESULTS WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. THERE IS NO INDICATION OF A PUMP MALFUNCTION. OCCLUSION ALARMS MAY HAVE EMITTED APPROPRIATELY DUE TO BLOCKAGE AND NOT A PUMP MALFUNCTION. THERE WERE NO ALARMS IN THE PUMP HISTORY LEADING UP TO THE HOSPITALIZATION, THEREFORE THE OCCLUSION ALARMS DID NOT CAUSE THE HOSPITALIZATION.

Description of Event or Problem · 1

THE PT'S MOTHER REPORTED THAT THE PT WAS ADMITTED TO THE HOSPITAL WITH A DIAGNOSIS OF DKA. THE PT HAD BEEN RECOVERING FROM PNEUMONIA FOR THE PAST MONTH PRIOR TO CALLING ANIMAS. THE PT'S MOTHER STATED THAT THE NURSING STAFF CHANGED OUT THE CARTRIDGE AND SITE AND THE PUMP CONTINUED TO EMIT OCCLUSION ALARMS. THE PT'S MOTHER MENTIONS A HISTORY OF OCCLUSION ALARMS WITH THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 16 YR