FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 1941309
·
Received December 16, 2010
Report
- Report Number
- 1720753-2010-04762
- Event Type
- Malfunction
- Date Received
- December 16, 2010
- Date of Event
- November 24, 2010
- Report Date
- December 16, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE AND USB ADAPTOR WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9900 SYSTEM HAD A HIGH CAPACITY DISK FAILURE ERROR AND WOULD NOT SAVE IMAGES. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|