FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS FLEX45

MDR report key: 1941098 · Received December 8, 2010

Report

Report Number
3005075853-2010-06912
Event Type
Malfunction
Date Received
December 8, 2010
Date of Event
September 13, 2010
Report Date
September 24, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). FIRING TRIGGER TEETH. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE ATW45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH TWO RELOADS PRESENT. RELOAD (A) WAS RECEIVED FULLY FIRED AND WITH CARTRIDGE LOCKOUT SPRING DAMAGED; RELOAD (B) WAS RECEIVED UNFIRED. THE DAMAGE TO THE RELOAD LOCKOUT SPRING IS CONSISTENT WITH DAMAGE OBSERVED WHEN ATTEMPTING TO FIRE AN ALREADY SPENT CARTRIDGE. THE RELOAD IS DESIGNED TO LOCK OUT AFTER A PARTIAL OR COMPLETE FIRE TO AVOID RE FIRING A PARTIALLY OR FULLY SPENT RELOAD. FIRING THROUGH THE LOCKOUT MECHANISM CAN BREAK THE DEVICE. FOR MORE INFO PLEASE REFER TO THE INSTRUCTIONS FOR USE. NO FUNCTIONAL TEST COULD BE PERFORMED WITH THE RETURNED INSTRUMENT DUE TO ITS CONDITION. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. HOWEVER, RELOAD (B) WAS TESTED FOR FUNCTIONALITY WITH A TEST INSTRUMENT AND IT FIRED, CUT, AND ALL THE STAPLES FORMED AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROTOMY PROCEDURE, THE DEVICE REMAINED BLOCKED AND COULD NOT WORK, THE DEVICE WOULD NOT FIRE. THE PROCEDURE WAS FINALLY PERFORMED WITH A COMPETITOR DEVICE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS FLEX45 GDW STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC NA G4RJ3W

Patients

Seq Age Sex Outcome Treatment
1 UNK