FDA Adverse Event Malfunction Summary report: N

LIFEPAK 15 MONITOR/DEFIBRILLATOR

MDR report key: 1941016 · Received December 8, 2010

Report

Report Number
3015876-2010-01344
Event Type
Malfunction
Date Received
December 8, 2010
Date of Event
November 8, 2010
Report Date
November 8, 2010
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K082937
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE TO POWER ON. PHYSIO REPLACED THE MAIN CONTROL KEYPAD ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD INTERMITTENTLY FAIL TO POWER ON WITH KNOWN GOOD BATTERIES. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 15 MONITOR/DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. 15 NA

Patients

Seq Age Sex Outcome Treatment
1 NA