FDA Adverse Event Malfunction Summary report: N

1/8" DRILL W/O STOP - 2/PKG

MDR report key: 1940998 · Received December 7, 2010

Report

Report Number
2249697-2010-01669
Event Type
Malfunction
Date Received
December 7, 2010
Date of Event
October 21, 2010
Report Date
November 15, 2010
Manufacturer
STRYKER ORTHOPAEDICS MAHWAH
Product Code
LXH
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: VISUAL INSPECTION INDICATED THE RETURNED DRILL HAS FRACTURED. THE FRACTURED POINT IS LOCATED AT THE TRANSITION POINT FROM SOLID BAR TO THE FLUTES. THE 1/8" DRILL W/O STOP IS MADE OF 17-4 PH STAINLESS STEEL WITH MINIMUM HARDNESS HRC 38. THE FRACTURE SURFACE WAS EXAMINED UNDER AN OPTICAL MICROSCOPE AT 20X MAGNIFICATION. THE EXAMINATION REVEALED THAT THE FRACTURE HAD OCCURRED AS A RESULT OF A BENDING OVERLOAD AS OPPOSED TO A TORSIONAL OVERLOAD, SUGGESTING USER MISUSE. THE INVESTIGATION CONCLUDED THAT THE DRILL WAS LIKELY SUBJECTED TO EXCESSIVE BENDING FORCES DUE TO EXCESSIVE ANGULATION DURING DRILLING. THE DRILL IS NOT INTENDED TO PERFORM UNDER EXCESSIVE ANGULATION AND LOADING DURING USE. THIS EVENT IS BELIEVED TO BE RELATED TO USER MISUSE.

Description of Event or Problem · 1

IT WAS REPORTED THAT, "DURING THE TKA, THE SURGEON WAS DRILLING THE PT TIBIA TO FIX A BASEPLATE, THE 1/8" DRILL WAS BROKEN. THE FRAGMENTS WAS REMOVED IMMEDIATELY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 1/8" DRILL W/O STOP - 2/PKG INSTRUMENT LXH STRYKER ORTHOPAEDICS MAHWAH NA UNK

Patients

Seq Age Sex Outcome Treatment
1 74 YR Other