FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 1940842 · Received December 17, 2010

Report

Report Number
2028159-2010-02392
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
November 17, 2010
Report Date
November 17, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

A COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND WAS ABLE TO REPLICATE THE REPORTED SYSTEM MESSAGE. THE SOLENOID SPACER WAS FOUND TO BE STUCK. THE SOLENOID BODY WAS CLEANED AND NEW SPACERS WERE INSTALLED. PREVENTIVE MAINTENANCE WAS ALSO PERFORMED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED RECEIVING A SYSTEM MESSAGE DURING A CASE. THE SYSTEM WAS SWITCHED OUT AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1