INFINITI VISION SYSTEM OZIL
MDR report key: 1940842
·
Received December 17, 2010
Report
- Report Number
- 2028159-2010-02392
- Event Type
- Malfunction
- Date Received
- December 17, 2010
- Date of Event
- November 17, 2010
- Report Date
- November 17, 2010
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
A COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND WAS ABLE TO REPLICATE THE REPORTED SYSTEM MESSAGE. THE SOLENOID SPACER WAS FOUND TO BE STUCK. THE SOLENOID BODY WAS CLEANED AND NEW SPACERS WERE INSTALLED. PREVENTIVE MAINTENANCE WAS ALSO PERFORMED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
A CUSTOMER REPORTED RECEIVING A SYSTEM MESSAGE DURING A CASE. THE SYSTEM WAS SWITCHED OUT AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT IMPACT REPORTED.