FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 1940757 · Received December 22, 2010

Report

Report Number
2953200-2010-02626
Event Type
Injury
Date Received
December 22, 2010
Date of Event
November 23, 2010
Report Date
November 23, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL RESULTS: (ENDOLEAK). (TYPE 2 ENDOLEAK).

Description of Event or Problem · 1

A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.7 CM ABDOMINAL AORTIC ANEURYSM. THE VESSELS WERE UNREMARKABLE. IT WAS REPORTED THAT AFTER THE CONTRALATERAL LIMB WAS IMPLANTED AN UNK TYPE OF ENDOLEAK WAS EVIDENT AT THE CONTRALATERAL GATE (SEE MFR REPORT # 2953200-2010-02624). ANOTHER CONTRALATERAL LIMB (SEE MFR REPORT # 2953200-2010-02625) WAS USED TO RELINE THE FIRST LIMB; HOWEVER, THE ENDOLEAK WAS STILL PRESENT. NO FURTHER INTERVENTION WAS DONE, AND THE DECISION WAS MADE TO MONITOR THE PT. THE PT HAD NOT BEEN EXCESSIVELY HEPARINIZED DURING THE PROCEDURE AND BALLOONING WAS NOT AGGRESSIVE. FILMS WERE EVALUATED BY AN INDEPENDENT PHYSICIAN AND HIS IMPRESSION IS AS FOLLOWS: THE INTRA-OPERATIVE ANGIO FINDINGS WERE SUGGESTIVE FOR A TYPE 3 ENDOLEAK FROM THE CONTRA GATE, BETWEEN THE TALENT BIFURCATED GRAFT AND THE FIRST CONTRALATERAL LIMB. THE ENDOLEAK DID NOT RESOLVE AFTER PLACEMENT OF THE ADD'L EXTENSION WITHIN THAT AREA. THE POST-OP CT SHOWS AN ENDOLEAK, BUT ONE THAT IS NOT VERY IMPRESSIVE, AND CLEARLY CONNECTED TO AT LEAST 1 PAIR OF PATENT AND BACK-FLOWING LUMBAR ARTERIES. THE PHYSICIAN COMMENTED, THIS WAS A TYPE 2 ENDOLEAK OR, AT A MINIMUM, A COMBINATION OF A TYPE 2 AND SOMETHING ELSE AND THAT THE PT SHOULD BE MONITORED. NO ADD'L CLINICAL SEQUELAE WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00446507

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention